C
SP-4430 ● 2-35
2-5
校正
Chapter 2 MEASUREMENT
2-5-1
校正の概要
Insert the Reagent Strips with the
same items, starting from the S1
channel on the Reagent Table.
If the number of calibration is set as
more than 3 for 2-point calibration,
calibration should be sperately
performed twice. In this case , set 3
Reagent Strips for the first
calibration.
(ex.: When the number of calibration
is 4, set 3 Reagent Strips for the first
calibration and 1 for the second .)
●
With Multi-type Reagent Strips,
1-point calibration cannot be
performed.
▲
●
Calibration with Multi-type Reagent Strips
Prepare Multi-type Reagent Strips.
Prepare twice as many Multi-type
Reagent Strips as the number of the
times calibration that will be
performed.
Set the Strip on the Reagent Table
from the right hand side.
●
●
Summary of Contents for SPOTCHEM EZ SP-4430
Page 1: ...b SPOTCHEM EZ SP 4430 Operating Manual ...
Page 8: ...C 07 SP 4430 ...
Page 9: ...b ...
Page 25: ...C 1 16 SP 4430 ...
Page 26: ...b ...
Page 31: ...e SP 4430 2 05 1 4 装置の設置 Chapter 2 MEASUREMENT 2 1 4 校正について The flowchart of calibration ...
Page 70: ...C 2 44 SP 4430 ...
Page 71: ...b ...
Page 74: ...C SP 4430 3 03 1 4 装置の設置 Chapter 3 SUB MENU 1 4 3 設設置後の始動と終了 ...
Page 97: ...C 3 26 SP 4430 ...
Page 98: ...b ...
Page 117: ...b ...
Page 129: ...b 5 12 SP 4430 Chapter 5 TROUBLESHOOTING ...
Page 130: ...b ...
Page 139: ...b 83 00139F ...