4-2
毎日のお手入れ
Chapter 4 MAINTENANCE
4-2-2
Cleaning the Tip Waste Case and Protective Cover
The Tip Waste Case holds used tips.
Its capacity is for 5 measurements
(※)
.Discard used tips and wash the Tip
Waste Case. Clean the Protective Cover if necessary.
(※)
A message suggesting the disposal of used tips is displayed every 5
measurements.
Slide the Centrifuge-equipped Multi Rack forward.
1.
Press [2] key on the MAIN MENU.
SUB MENU 1/2 is displayed.
Press [3] key .
The Table Cover opens and the
Reagent Table and the Centrifuge-
equipped Multi Rack slide forward.
Turn OFF the power.
●
●
●
Discard Tips.
2.
Remove the Tip Waste Case from
the Centrifuge-equipped Multi Rack.
Discard tips.
●
●
Requirements
Alcohol, Cloth and Protective gloves
Wear protective gloves to prevent exposure to pathogenic microbes.
WARNING
Discard used samples, tips and protective gloves separately from general
waste according to local regulations on biohazardous waste.
CAUTION
1 . R e s u l t s 2 . P A R AM
3 . M a i n t e n a n c e
( 1/2 )
1 . M e a s u r e 2 . S u b m e n u
3 . C a l i b r a t e
( 1/1 )
R e a d y .
P l e a s e
t u r n o f f .
Summary of Contents for SPOTCHEM EZ SP-4430
Page 1: ...b SPOTCHEM EZ SP 4430 Operating Manual ...
Page 8: ...C 07 SP 4430 ...
Page 9: ...b ...
Page 25: ...C 1 16 SP 4430 ...
Page 26: ...b ...
Page 31: ...e SP 4430 2 05 1 4 装置の設置 Chapter 2 MEASUREMENT 2 1 4 校正について The flowchart of calibration ...
Page 70: ...C 2 44 SP 4430 ...
Page 71: ...b ...
Page 74: ...C SP 4430 3 03 1 4 装置の設置 Chapter 3 SUB MENU 1 4 3 設設置後の始動と終了 ...
Page 97: ...C 3 26 SP 4430 ...
Page 98: ...b ...
Page 117: ...b ...
Page 129: ...b 5 12 SP 4430 Chapter 5 TROUBLESHOOTING ...
Page 130: ...b ...
Page 139: ...b 83 00139F ...