Chapter 2 MEASUREMENT
2-5-3
キャリブレータ校正
2-34 ● SP-4430
Set the Tip.
5.
Set the tips in the tip hole of the
Centrifuge-equipped Multi Rack
using tweezers.
●
Make sure that the Tip Waste Case
is set in position.
●
When performing 1-point
calibration (High), proceed to
procedure 9.
▲
Set the calibrator “Low”.
6.
Remove the caps and set the
calibrator “Low” to the port of the
Centrifuge-equipped Multi Rack.
●
Set the Reagent Strips.
7.
●
Calibration with Single-type Reagent Strips
Prepare Single-type Reagent Strips.
When performing 2-point calibration,
prepare twice as many Reagent
Strips as the number of times
calibration that will be performed.
For 1-point calibration, prepare the
same number of times calibration
that will be performed.
●
××
×
××
×
Do NOT touch the end of the Tip
with bear hands. If the tips are
soiled, correct sampling may not
be done and this will lead to
incorrect calibration results.
IMPORTANT
NOTE
Be sure to remove the caps before
calibration to avoid damage to the
nozzle.
Be sure to insert the tip of the strip in the groove of the reagent table so that
the Reagent Strip stays firmly in place. If the strip is placed out of the groove,
it may cause jam or correct measurement results may not be obtained.
IMPORTANT
NOTE
Make sure that the protective cover
is closed before the measurement.
If the protective cover remains
open, it may cause a trouble.
Summary of Contents for SPOTCHEM EZ SP-4430
Page 1: ...b SPOTCHEM EZ SP 4430 Operating Manual ...
Page 8: ...C 07 SP 4430 ...
Page 9: ...b ...
Page 25: ...C 1 16 SP 4430 ...
Page 26: ...b ...
Page 31: ...e SP 4430 2 05 1 4 装置の設置 Chapter 2 MEASUREMENT 2 1 4 校正について The flowchart of calibration ...
Page 70: ...C 2 44 SP 4430 ...
Page 71: ...b ...
Page 74: ...C SP 4430 3 03 1 4 装置の設置 Chapter 3 SUB MENU 1 4 3 設設置後の始動と終了 ...
Page 97: ...C 3 26 SP 4430 ...
Page 98: ...b ...
Page 117: ...b ...
Page 129: ...b 5 12 SP 4430 Chapter 5 TROUBLESHOOTING ...
Page 130: ...b ...
Page 139: ...b 83 00139F ...