Chapter 2 MEASUREMENT
2-5-3
Calibration by calibrator kit
Use the SPOTCHEM Calibrator Kit for calibration. For Hb items, use the
SPOTCHEM Calibrator Hb Kit.
Display the calibration menu.
2.
Press [3] key on the MAIN MENU.
CALIBRATION MENU is displayed.
●
“CARD” means calibration by
magnetic calibration card and
“Cal.” means calibration by
calibrator kit.
▲
Set the calibration type by calibrator kit.
1.
When changing the calibration type from calibration by magnetic card (CARD) to
the calibration by calibrator kit (Cal.), change the setting by referring to “see 3-3-2
Enter parameters”).
If the calibration type is already set for calibrator kit, this operation is not necessary.
●
Prepare the reagent strip and calibrator.
3.
Prepare the required amount of
Reagent Strips according to the
following table.
●
When performing the
calibration by calibrator kit
using Multi Reagent Strips,
calibration mode of each
measurement item (single item)
attached to Multi Reagent Strip
is used.
▲
Wear protective gloves to prevent exposure to pathogenic microbes.
WARNING
Discard used samples, tips and protective gloves separately from general
waste according to local regulations on biohazardous waste.
CAUTION
Be careful not to spill sample blood when uncapping the tube.
CAUTION
1 . C A R D
2 . C a l .
3 . I n f o r m a t i o n
( 1/1 )
1 . M e a s u r e 2 . S u b m e n u
3 . C a l i b r a t e
( 1/1 )
Summary of Contents for SPOTCHEM EZ SP-4430
Page 1: ...b SPOTCHEM EZ SP 4430 Operating Manual ...
Page 8: ...C 07 SP 4430 ...
Page 9: ...b ...
Page 25: ...C 1 16 SP 4430 ...
Page 26: ...b ...
Page 31: ...e SP 4430 2 05 1 4 装置の設置 Chapter 2 MEASUREMENT 2 1 4 校正について The flowchart of calibration ...
Page 70: ...C 2 44 SP 4430 ...
Page 71: ...b ...
Page 74: ...C SP 4430 3 03 1 4 装置の設置 Chapter 3 SUB MENU 1 4 3 設設置後の始動と終了 ...
Page 97: ...C 3 26 SP 4430 ...
Page 98: ...b ...
Page 117: ...b ...
Page 129: ...b 5 12 SP 4430 Chapter 5 TROUBLESHOOTING ...
Page 130: ...b ...
Page 139: ...b 83 00139F ...