f
SP-4430 ● 2-07
1-4
装置の設置
Chapter 2 MEASUREMENT
2-2-2
Handling samples
Handling samples slightly differs
depending on test items. Follow the
instructions on the package insert of
SPOTCHEM Reagent Strips.
Wear
protective gloves
Blood is used as a measurement sample
with this analyzer. Be aware that blood may
be contaminated with pathogenic microbes
that may cause infectious disease. Exercise
utmost care whenever handling blood.
Incorrect or imprecise procedures may
result in exposure to pathogenic microbes.
2-2-3
Handling Reagent Strips
●
●
Do NOT use expired Reagent Strips.
Do NOT use deteriorated strips of which the
reagent fields show any sign of discoloration
or deformation, even if they are before the
e x p i r a t i o n d a t e . O t h e r w i s e , c o r r e c t
measurement results may not be obtained.
●
Take Reagent Strips out of the refrigerator
20 minutes before measurement and allow
them to reach room temperature (10 to
30
℃
). Otherwise, correct measurement
results may not be obtained.
●
Open a reagent strip package immediately
before use. If the opened reagent strip is
left for a while, it will absorb water from the
air or gather dust, which may result in
incorrect measurement results.
●
Do NOT touch the reagent field on the
strip with fingers. Sebum on the
reagent field may result in incorrect
measurement results.
●
When a new reagent strip box is
opened, perform calibration by magnetic
card using the supplied Reagent Card
(see 2-5-2).
●
Summary of Contents for SPOTCHEM EZ SP-4430
Page 1: ...b SPOTCHEM EZ SP 4430 Operating Manual ...
Page 8: ...C 07 SP 4430 ...
Page 9: ...b ...
Page 25: ...C 1 16 SP 4430 ...
Page 26: ...b ...
Page 31: ...e SP 4430 2 05 1 4 装置の設置 Chapter 2 MEASUREMENT 2 1 4 校正について The flowchart of calibration ...
Page 70: ...C 2 44 SP 4430 ...
Page 71: ...b ...
Page 74: ...C SP 4430 3 03 1 4 装置の設置 Chapter 3 SUB MENU 1 4 3 設設置後の始動と終了 ...
Page 97: ...C 3 26 SP 4430 ...
Page 98: ...b ...
Page 117: ...b ...
Page 129: ...b 5 12 SP 4430 Chapter 5 TROUBLESHOOTING ...
Page 130: ...b ...
Page 139: ...b 83 00139F ...