Chapter 2 MEASUREMENT
2-3-3
Checks before measurement
Thermal printer paper.
1.
If a red line appears on both sides of
the printer paper, replace it with a
new roll (see“4-3-1 Thermal printer
paper replacement”).
●
Reagent Table.
2.
Check the Reagent Table and clean
it if necessary (see “4-2-1 Cleaning
the Reagent Table”).
●
Centrifuge-equipped Multi Rack.
3.
Make sure that the Centrifuge-
equipped Multi Rack has no used
samples in it. If any, discard them.
Install the Tip Waste Case.
●
●
Check each part before performing Normal Measurement or Calibration.
Setting the parameters.
4.
Set the parameters as necessary. If the same parameter settings are continuously
used, resetting is not necessary. See “Chapter 3 SUB MENU” for details.
●
Wear protective gloves to prevent exposure to pathogenic microbes.
WARNING
Discard used samples, tips and protective gloves separately from general
waste according to the local regulations on biohazardous waste.
CAUTION
To check the present parameter settings, print them out (see “3-3-1 Printing
parameters”).
IMPORTANT
Summary of Contents for SPOTCHEM EZ SP-4430
Page 1: ...b SPOTCHEM EZ SP 4430 Operating Manual ...
Page 8: ...C 07 SP 4430 ...
Page 9: ...b ...
Page 25: ...C 1 16 SP 4430 ...
Page 26: ...b ...
Page 31: ...e SP 4430 2 05 1 4 装置の設置 Chapter 2 MEASUREMENT 2 1 4 校正について The flowchart of calibration ...
Page 70: ...C 2 44 SP 4430 ...
Page 71: ...b ...
Page 74: ...C SP 4430 3 03 1 4 装置の設置 Chapter 3 SUB MENU 1 4 3 設設置後の始動と終了 ...
Page 97: ...C 3 26 SP 4430 ...
Page 98: ...b ...
Page 117: ...b ...
Page 129: ...b 5 12 SP 4430 Chapter 5 TROUBLESHOOTING ...
Page 130: ...b ...
Page 139: ...b 83 00139F ...