74
Information screens
Frontpage w. status
Airtec® ROinline
Software V. 3.00
Status:
OK
Time for Service
Ext. Stop Active
Startup screen - with software version no. and status.
Status:
OK or ALARM.
Time for Service: Shown when it is time for service once a year .
Ext. Stop Active:
Shown if external stop is activated.
Pressure & flow
Vessel Pressure
Bar 4.92
Prod. l/h 300
Drain l/h 96
Drain % 32
Information about the pressure in the vessel and Flow.
Bar
Prod. l/h
Drain l/h
Drain %
Vessel pressure in Bar.
Production flow in l/h.
Drain flow in l/h.
Drain flow in %.
Showing the actual flow during the production period, when the
pump stops the start values will be displayed.
Pump, valves & sensors
Valve In:
Closed
Water In:
OK
Pump:
Stopped
Hygiene:
Closed
Prod. Flow:
No
Drain Flow:
No
Information about status on pump, sensors and valves
Valve In:
Water In:
Pump:
Hygiene:
Prod. Flow:
Drain Flow:
Open or Closed
OK or Fault
Running or Stopped
Open or Closed
Yes or No
Yes or No
Password
Type Password
to enter setup
Password:
1300
Page to access user setup menu, password protected.
Password
User setup:
1300.
Maintenance: 9963.
Flow calibration: 3112.
When typing the password, the user setup menu will be available
for 10 min. then the menu will be hidden again, this is to prevent
unauthorized access.
1/9 Inputs
I:
1/9
1
2
3456
7
89
10+ 0123456789
20+ 01234
Information on ditial input - screen 1/9.
If an Input is ”ON” the number of the Input gets black
(as shown with no. 2 and 7).
2/9 Outputs
Q:
2/9
1
2
3
456789
10+ 0123456789
20+ 01234
Information on output - screen 2/9.
If an output is ”ON” the number of the output gets black
(as shown with no. 1 and 3).
3/9 Analog inputs
AI:
3/9
1: 000000
2: 000489
3: 000000
4: 000000
5: 000000
Information on analog input - screen 3/9.
On this page you can read the analog input values.
The only analog input used in the RO Inline™ is no. 2, used for the
pressure transmitter 0 - 10 bars/0-10 V.
In this example the display shows 00489 which is 4,89 bars.
9/9 Arrow screen
ESC+C.. 9/9
~
|
}
Arrow screen 9/9.
[ESC]
+
[ESC]
+
~
Reset all alarms.
Manual start hygiene rinse.
Shared user and maintenance setup screens
Hours & power up
-Hours-
System
XXXXXX
Pump
XXXXXX
S.Service
XXXXXX
-Log-
PowerUp XXXXXX
Information about operating hours and power up log.
Syst.
Pump
S.Service
PowerUp
Total system hours.
Pump operating hours.
Hours since last service.
Shows how many times the system power has been
turned on.
Edit hours eg. if the pump is changed or if the system has been
updated with new software.
Only avalible in maintenance (Pwd: 9963).
Alarm log
No Water XXXXXX
External XXXXXX
Temp.Motor XXXXXX
Low Prod. XXXXXX
Low Drain XXXXXX
Fault K1 XXXXXX
Alarm log, for:
No Water
External
Temp.Motor
Low Prod.
Low Drain
Fault K1
Low water pressure.
External alarm.
Thermal protection of the motor.
Low production flow.
Low drain flow.
Fault on contactor K1.
Pressure setup
-Pressure Setup
Stop (3-6)
5
Hyst.(1-4)
3
Start Bar: 2
**Setup out**
of Range
Vessel pressure setup:
Stop
Hyst.
Start
**Setup...
Pressure to stop production - min. 3, max. 6 bar.
Hysteresis for starting production 1-4bar.
Production start pressure.
Setup is out of range, will reset to default.
The pressure range in the vessel is from 2 to 6 bar.
Flow setup
-Production Set
Min. l/h
200
Delay Sec.
20
-Drain Setup
Min. %
25
Delay Sec.
20
Flow setup:
Min. l/h
Delay Sec.
Min. %
Delay Sec.
The min. continuous flow needed, l/h.
Delay before the alarm, sec.
The min. continuous flow to the drain, %
Delay before the alarm, sec.
On Prod. alarm, will the RO Inline™ continue to produce permeate.
Maintenance setup screens
Service
Service Info.
Show
1
Reset = Set to 1
Service
0
Log.
0
Settings: Service and resetting service and log.
Show
Service
Log.
Sets if time for service message is shown.
0 = Hide; 1 = Show.
Set to 1 to reset service counter.
Set to 1 to reset alarm log.
Alarm in/out
Set Alarm In/Out
IN
NO
1
Delay Sec.
15
W. as
Alarm
0
Out
NC
0
Settings: Alarm in/out NC/NO and delay.
IN
Delay
W. as Alarm
Out
Alarm in, 0 = NC; 1 = NO.
Alarm in delay in sec. before the system stops.
Ext. alarm In work as 0 = alarm; 1 = stop.
if stop then the system will start automaticly.
Alarm out, 0 = NC; 1 = NO.
Hygiene & delay
Hours between
Hyg. Rinse
6
Delay in Sec.
Pump Start
5
No Water
15
Settings: Hygiene and delays.
Hyg. Rinse
Pump Start
No Water
Hours between hygiene rinse.
Delay between water in opens and the pump start.
Delay before no water alarm.
201218