Electrical installation
104
GasEye Cross Duct SG and MG
Operating instructions, 05/2021
PROCESS.PRESS_MANUAL_VALUE 0x05
5 16 -bits of floating point value (first
part)
Process pressure manual value in
0x06
6 16 -bits of floating point value
(second part)
Process pressure manual value in
MEAS.PATH_LENGTH_CH1
0x07
7 16 -bits of floating point value (first
part)
Measuring path length channel 1
0x08
8 16 -bits of floating point value
(second part)
Measuring path length channel 1
MEAS.PATH_LENGTH_CH2
0x09
9 16 -bits of floating point value (first
part)
Measuring path length channel 2
0x0A
10 16 -bits of floating point value
(second part)
Measuring path length channel 2
MEAS.RESPONSE_TIME_T90
0x0B
11 16 -bits of floating point value (first
part)
Response time (T90) – const for IIR
filter
0x0C
12 16 -bits of floating point value
(second part)
Response time (T90) – const for IIR
filter
GAS101.SPAN_CALIBRATION
0x0D
13 16 -bits of floating point value (first
part)
Span calibration factor
0x0E
14 16 -bits of floating point value
(second part)
Span calibration factor
GAS101.OFFSET
0x0F
15 16 -bits of floating point value (first
part)
Span offset value
0x10
16 16 -bits of floating point value
(second part)
Span offset value
GAS102.SPAN_CALIBRATION
0x11
17 16 -bits of floating point value (first
part)
Span calibration factor
0x12
18 16 -bits of floating point value
(second part)
Span calibration factor
GAS102.OFFSET
0x13
19 16 -bits of floating point value (first
part)
Span offset value