Purging (non-ATEX)
198
GasEye Cross Duct SG and MG
Operating instructions, 05/2021
Tube and process purging
If it is necessary to purge the tubes and process side of the instrument the user should
adapt the purging interfaces.
1. Screw in 1/8 inch fittings into points indicated by red arrows.
Figure 194. Tube and process purging.
The following versions of purge fittings exist:
Fitting type
Thread
Material
Picture
“Insert fitting”
for
¼”
tube
1/8”
BSPT
thread
SS/PTFE
“Insert fitting”
for
6mm
tube
SS/PTFE
¼”
compression
fitting
SS316
6mm
double
compression
fitting
SS316
Table 56. Possible purge fittings versions for GasEye Cross Duct tube or
alignment flange.
Tube purge inlet
Tube purge outlet
Process purge