![Airoptic GasEye User Manual Download Page 91](http://html1.mh-extra.com/html/airoptic/gaseye/gaseye_user-manual_2878696091.webp)
Electrical installation
GasEye Cross Duct
Operating instructions, 05/2021
91
GAS102.OFFSET
0x91
145 16 -bits of floating point value (first
part)
Span offset value
0x92
146 16 -bits of floating point value
(second part)
Span offset value
GAS103.SPAN_CALIBRATION
0x93
147 16 -bits of floating point value (first
part)
Span calibration factor
0x94
148 16 -bits of floating point value
(second part)
Span calibration factor
GAS103.OFFSET
0x95
149 16 -bits of floating point value (first
part)
Span offset value
0x96
150 16 -bits of floating point value
(second part)
Span offset value
GAS104.SPAN_CALIBRATION
0x97
151 16 -bits of floating point value (first
part)
Span calibration factor
0x98
152 16 -bits of floating point value
(second part)
Span calibration factor
GAS104.OFFSET
0x99
153 16 -bits of floating point value (first
part)
Span offset value
0x9A
154 16 -bits of floating point value
(second part)
Span offset value
GAS105.SPAN_CALIBRATION
0x9B
155 16 -bits of floating point value (first
part)
Span calibration factor
0x9C
156 16 -bits of floating point value
(second part)
Span calibration factor
GAS105.OFFSET
0x9D
157 16 -bits of floating point value (first
part)
Span offset value
0x9E
158 16 -bits of floating point value
(second part)
Span offset value
GAS106.SPAN_CALIBRATION
0x9F
159 16 -bits of floating point value (first
part)
Span calibration factor
0xA0
160 16 -bits of floating point value
(second part)