Installation
38
GasEye Cross Duct SG and MG
Operating instructions, 05/2021
11. There should be a light spot
visible on the fine aligning
unit screen. By sliding the
fine aligning unit, the bright
spot may be focused on the
screen. Find a position at
which the spot becomes
the smallest and sharpest
and tighten the screw in the
aiming tool.
12. Again, use the provided
Allen key to align the light
spot in the desired point on
the crosshair. The optimal
aligning point can be found
on the sticker under the
analyzer
lid
(both
transmitter/central unit or
receiver – see picture
below)
Figure 33. Finding the optimal aligning
point.
Figure 34. Target aligning point sticker (on the bottom of receiver unit lid).
CORRECT
INCORRECT