Electrical installation
GasEye Cross Duct
Operating instructions, 05/2021
93
AOUT.FORCE_MANUAL_MODE
ENABLE
0xB3
179 Force manual mode for all analog
outputs
AOUT.SCALE_ENALE
0xB4
180 Enable scaling range for all analog
outputs
AOUT.CALIBRATED
0xB5
181 Reserved - future use
AOUT1.SELECT_SIGNAL
0xB6
182 Measurement signal selection for
the output
AOUT1.MANUAL_VALUE
0xB7
183 16 -bits of floating point value (first
part)
Manual mode value in scaling range
0xB8
184 16 -bits of floating point value
(second part)
Manual mode value in scaling range
AOUT1.SCALE_MIN
0xB9
185 16 -bits of floating point value (first
part)
Scaling range value corresponds to 4
mA
0xBA
186 16 -bits of floating point value
(second part)
Scaling range value corresponds to 4
mA
AOUT1.SCALE_MAX
0xBB
187 16 -bits of floating point value (first
part)
Scaling range value corresponds to
20 mA
0xBC
188 16 -bits of floating point value
(second part)
Scaling range value corresponds to
20 mA
AOUT1.A
0xBD
189 16 -bits of floating point value (first
part)
Reserved - future use
0xBE
190 16 -bits of floating point value
(second part)
Reserved - future use
AOUT1.B
0xBF
191 16 -bits of floating point value (first
part)
Reserved - future use
0xC0
192 16 -bits of floating point value
(second part)
Reserved - future use
AOUT2.SELECT_SIGNAL
0xC1
193 Measurement signal selection for
the output
AOUT2.MANUAL_VALUE
0xC2
194 16 -bits of floating point value (first
part)
Manual mode value in scaling range