Electrical installation
92
GasEye Cross Duct SG and MG
Operating instructions, 05/2021
Span calibration factor
GAS106.OFFSET
0xA1
161 16 -bits of floating point value (first
part)
Span offset value
0xA2
162 16 -bits of floating point value
(second part)
Span offset value
GAS107.SPAN_CALIBRATION
0xA3
163 16 -bits of floating point value (first
part)
Span calibration factor
0xA4
164 16 -bits of floating point value
(second part)
Span calibration factor
GAS107.OFFSET
0xA5
165 16 -bits of floating point value (first
part)
Span offset value
0xA6
166 16 -bits of floating point value
(second part)
Span offset value
GAS108.SPAN_CALIBRATION
0xA7
167 16 -bits of floating point value (first
part)
Span calibration factor
0xA8
168 16 -bits of floating point value
(second part)
Span calibration factor
GAS108.OFFSET
0xA9
169 16 -bits of floating point value (first
part)
Span offset value
0xAA
170 16 -bits of floating point value
(second part)
Span offset value
TEMP.SPAN_CALIBRATION
0xAB
171 16 -bits of floating point value (first
part)
Span calibration factor
0xAC
172 16 -bits of floating point value
(second part)
Span calibration factor
TEMP.OFFSET
0xAD
173 16 -bits of floating point value (first
part)
Span offset value
0xAE
174 16 -bits of floating point value
(second part)
Span offset value
DOUT.DO1
0xAF
175 Signal selection for digital output
DOUT.DO2
0xB0
176 Signal selection for digital output
DOUT.DO3
0xB1
177 Signal selection for digital output
DOUT.DO4
0xB2
178 Signal selection for digital output