Installation
22
GasEye Cross Duct SG and MG
Operating instructions, 05/2021
Installation
Safety information
GasEye Cross Duct is classified as a Class 1 laser product. The emitted laser
radiation is invisible (near infrared) and not hazardous to an unprotected eye. GasEye
Cross Duct has warning labels placed at positions specified in the EN 60825-1:2014-
11 norm.
NOTICE
Internal invisible IR laser is a
Class I
product.
WARNING
Heat safety
Some metal parts and piping placed near the sensors might be at elevated
temperatures due to high temperature purging - either from steam or from air. To
avoid severe burns these parts must either be isolated or equipped with protective
metal sheets. Always use protective gloves in the vicinity of such hot parts.
WARNING
Pressure safety
It is possible to mount the GasEye Cross Duct to processes with elevated pressure.
Although the window of the process interface is burst-tested up to 4 MPa (40 bar),
the customer should define a safety factor appropriate to his application.
WARNING
Electrical safety
Mounting or unmounting of the instrument must be performed with the power cord
DISCONNECTED.
The power supply must be connected only when the instrument
is fully adjusted and secured.