3-9
Function Reference
Amplitude Units (Amplitude Setup Panel)
When AMPCOR is turned on, the correction points are applied across the
active measurement range and added to all measurement results. Between
points, the correction values are interpolated linearly or logarithmically. When
measuring at wavelengths outside the first and last correction points, the first
or last value (as appropriate) is used as the correction value.
Whenever AMPCOR is active, the currently selected correction set is dis-
played in the lower left corner of the screen. For example, if correction set
number one is selected, “A1” is displayed.
Refer to “Amplitude Correction
for an overview of the amplitude correction
remote commands.
Key Path
Amplitude > Amplitude Setup...
Related Functions
Reference Level, Sensitivity
Remote
Commands
(Amplitude Display Mode):DISPlay:WINDow:TRACe:Y:SCALe LINear
DISPlay:WINDow:TRACe:Y:SCALe LOGarithmic
(Amplitude Units):UNIT:POWer
(Auto Zero):CALibration:ZERO
(Reference Level Position):DISPlay:WINDow:TRACe:Y:SCALe:RPOSition
Amplitude Units (Amplitude Setup Panel)
Selects the amplitude scale’s units. Selecting Auto for amplitude units allows
power to be shown in dB when using the logarithmic scale and in watts when
using the linear scale. Selecting W forces the readout to be in watts, regardless
of the amplitude scaling.
Key Path
Amplitude > Amplitude Setup > Amplitude Units
Related Functions
Display Mode, Trace Marker, Scale/Div
Remote
Commands
UNIT:POWer
Summary of Contents for 86140B Series
Page 2: ...Agilent 86140B Series Optical Spectrum Analyzer User s Guide ...
Page 11: ...1 4 Getting Started Product Overview Agilent 86140B Front and Rear Panels ...
Page 12: ...1 5 Getting Started Product Overview ...
Page 14: ...1 7 Getting Started Product Overview Figure 1 2 Display Annotations ...
Page 28: ...1 21 Getting Started The Softkey Panels ...
Page 33: ...1 26 Getting Started The Softkey Panels The Systems Menus continued ...
Page 41: ...1 34 Getting Started Product Options and Accessories ...
Page 79: ...2 38 Using the Instrument Analyzer Operating Modes ...
Page 80: ...3 Function Reference ...
Page 239: ...6 2 Maintenance Changing the Printer Paper Changing the Printer Paper ...