![Agilent Technologies 86140B Series User Manual Download Page 211](http://html.mh-extra.com/html/agilent-technologies/86140b-series/86140b-series_user-manual_2867952211.webp)
5-12
Status Listings
SCPI-Defined Errors
–222 desc
=
“Data out of range”
help = “A numeric value was entered which is outside the legal range of values
for the parameter. The name of the parameter is listed at the end of the error
message.”
[Indicates that a legal program data element was parsed but could not be
executed because the interpreted value was outside the legal range as defined
by the device (see IEEE 488.2, 11.5.1.1.5.)]
–222 desc
=
“Span out of range”
help = “A numeric value was entered which is outside the legal range of values
for the span setting. This occurs if the value is too large less than zero or
between zero and 0.2nm. Zero span may not be entered by using the DOWN
arrow key from 0.2nm; a numeric value of zero must be entered instead.”
[Indicates that a legal program data element was parsed but could not be
executed because the interpreted value was outside the legal range as defined
by the device (see IEEE 488.2, 11.5.1.1.5.)]
–223 desc
=
“Too much data”
help = ““
[Indicates that a legal program data element of block, expression, or string type
was received that contained more data than the device could handle due to
memory or related device-specific requirements.]
–224 desc
=
“Illegal parameter value”
help = ““
[Used where exact value, from a list of possibles, was expected.]
–257 desc
=
“File name error”
help = ““
[Indicates that a legal program command or query could not be executed
because the file name on the device media was in error; for example, an attempt
was made to copy to a duplicate file name. The definition of what constitutes a
file name error is device-specific.]
–272 desc
=
“Macro Exec Error”
help = ““
[Indicates that a syntactically legal macro program data sequence could not be
executed due to some error in the macro definition (see IEEE 488.2, 10.7.6.3.)]
Table 5-2. Execution Errors (2 of 3)
Error Number
Error Description [description/explanation/examples]
Summary of Contents for 86140B Series
Page 2: ...Agilent 86140B Series Optical Spectrum Analyzer User s Guide ...
Page 11: ...1 4 Getting Started Product Overview Agilent 86140B Front and Rear Panels ...
Page 12: ...1 5 Getting Started Product Overview ...
Page 14: ...1 7 Getting Started Product Overview Figure 1 2 Display Annotations ...
Page 28: ...1 21 Getting Started The Softkey Panels ...
Page 33: ...1 26 Getting Started The Softkey Panels The Systems Menus continued ...
Page 41: ...1 34 Getting Started Product Options and Accessories ...
Page 79: ...2 38 Using the Instrument Analyzer Operating Modes ...
Page 80: ...3 Function Reference ...
Page 239: ...6 2 Maintenance Changing the Printer Paper Changing the Printer Paper ...