5-23
Status Listings
OSA Warnings
5062 desc
=
“ADC Triggered Sweep Too Fast”
help = “A sweep was taken with one of the ADC trigger modes enabled. The
sweep rate was too fast to allow trace data to be acquired for every
wavelength. To resolve this problem increase sweep time or increase the
frequency the external trigger input signal.”
5063 desc
=
“Key disabled during applications”
help = “The key you pressed is not active while an application is running. Exiting
the application should re-enable the key.”
5064 desc
=
“Invalid measurement file”
help = “An attempt was made to recall a measurement file which is contains
invalid or corrupt data.”
5065 desc
=
“Error detected in DSP sub-system”
help = “The Digital Signal Processor has reported an unexpected error. Please
record the hexadecimal number listed with the error and cycle power. If the error
persists, contact the Agilent support center nearest your location.”
5066
desc = “Error in Enhanced Wavelength Calibration”
help = “Enhanced Wavelength Calibration failed. Check OSA system. If the error
persists, contact the Agilent Service Center.
5067
desc = “OSNR marker not allowed.”
help = “A request was made to enable the OSNR marker readout. The readout
cannot be enabled for one of the following reasons: OSNR markers cannot be
used on traces which have X axis units other than meters. This typically occurs
in zero span where the X axis units are seconds. OSNR marks cannot be used on
traces which have Y axis units other than power. this typically happens when
trace math is on and the trace Y axis represents a ratio.”
5068
desc = “The configured SHARE does not exist.”
help = “A request was made to a remote file or printer share that does not exist
or cannot be found.”
5069
desc = “The configured SHARE cannot be accessed.”
help = “A request was made to a remote file or printer share that cannot be
accessed using the configured USER, PASSWORD, or DOMAIN/WORKGROUP.”
5070
desc = “Trajectory add failed.”
help = “The trajectory add procedure failed. This usually occurs when the
trajectory table is full or the computed trajectory table is invalid. The correct this
problem, try AUTO ALIGN PRESET and the AUTO ALIGN & ADD TO TRAJECTORY
procedure again. If the error persists, contact the Agilent Service Center.
Table 5-6. OSA Warnings (7 of 12)
Error Number
Error Description [description/explanation/examples]
Summary of Contents for 86140B Series
Page 2: ...Agilent 86140B Series Optical Spectrum Analyzer User s Guide ...
Page 11: ...1 4 Getting Started Product Overview Agilent 86140B Front and Rear Panels ...
Page 12: ...1 5 Getting Started Product Overview ...
Page 14: ...1 7 Getting Started Product Overview Figure 1 2 Display Annotations ...
Page 28: ...1 21 Getting Started The Softkey Panels ...
Page 33: ...1 26 Getting Started The Softkey Panels The Systems Menus continued ...
Page 41: ...1 34 Getting Started Product Options and Accessories ...
Page 79: ...2 38 Using the Instrument Analyzer Operating Modes ...
Page 80: ...3 Function Reference ...
Page 239: ...6 2 Maintenance Changing the Printer Paper Changing the Printer Paper ...