7-4
Specifications and Regulatory Information
Definition of Terms
Amplitude
Scale Fidelity
refers to the potential errors in amplitude readout at ampli-
tudes other than at the calibration point. This specification is sometimes
called linearity.
Flatness
defines a floating band which describes the error in signal amplitude
over the indicated wavelength range. (This error may be removed at a given
wavelength by performing the user amplitude calibration.)
Polarization Dependence
refers to the amplitude change that can be seen by
varying the polarization of the light entering the OSA. This is not to be con-
fused with amplitude variations caused by the varying distribution of energy
between the different modes in fiber that is multimode at the wavelength of
interest.
Sensitivity
Sensitivity
is defined as the signal level that is equal to six times the RMS
value of the noise. Displayed sensitivity values are nominal. Slightly lower val-
ues may have to be entered to achieve specified sensitivity.
Dynamic range
Dynamic Range
is a measure of the ability to see low-level signals that are
located very close (in wavelength) to a stronger signal. In electrical spectrum
analyzers, this characteristic is generally called shape factor.
Sweep time
Maximum Sweep Rate
refers to the maximum rate that the instrument is
able to acquire data and display it. This rate may be limited by multiple inter-
nal processes when using default number of trace points.
Sweep Cycle Time
refers to the time required to make a complete sweep and
prepare for the next sweep. It can be measured as the time from the start of
one sweep to the start of the next sweep.
Summary of Contents for 86140B Series
Page 2: ...Agilent 86140B Series Optical Spectrum Analyzer User s Guide ...
Page 11: ...1 4 Getting Started Product Overview Agilent 86140B Front and Rear Panels ...
Page 12: ...1 5 Getting Started Product Overview ...
Page 14: ...1 7 Getting Started Product Overview Figure 1 2 Display Annotations ...
Page 28: ...1 21 Getting Started The Softkey Panels ...
Page 33: ...1 26 Getting Started The Softkey Panels The Systems Menus continued ...
Page 41: ...1 34 Getting Started Product Options and Accessories ...
Page 79: ...2 38 Using the Instrument Analyzer Operating Modes ...
Page 80: ...3 Function Reference ...
Page 239: ...6 2 Maintenance Changing the Printer Paper Changing the Printer Paper ...