2-14
Using the Instrument
Calibrating Wavelength Measurements
Calibrating Wavelength Measurements
Environmental variations such as air pressure, temperature, and humidity can
affect the index of refraction of air in the monochromator of the optical spec-
trum analyzer (OSA). This section discusses calibration methods that you can
use to improve the wavelength accuracy in the Agilent 8614X OSA’s.
and to
“Calibrator Multi-Pt Align” on page 3-18
N o t e
Many aspects of remotely programming the optical spectrum analyzers are discussed
in Product Note 86140-2R, Wavelength Calibration for the 86140X Series Optical
Spectrum Analyzers (Literature part number 5980-0043E).
Overview
Wavelength calibration routines improve wavelength accuracy by determining
errors and correcting them with offsets, using linear interpolation when neces-
sary. For maximum wavelength accuracy, calibration points spaced a maxi-
mum of 10 nm apart are recommended.
You can perform a wavelength calibration by using one of the following meth-
ods:
• Manual Method using Internal Calibrator
• Remote Method using Internal Calibrator
• Manual Method using an External Single Wavelength Source
• Remote Method using an External Single Wavelength Source
• External Multipoint Wavelength Calibration
These calibration routines should only be performed after the instrument’s
temperature has been stabilized by a minimum of 1 hour of continuous opera-
tion.
Summary of Contents for 86140B Series
Page 2: ...Agilent 86140B Series Optical Spectrum Analyzer User s Guide ...
Page 11: ...1 4 Getting Started Product Overview Agilent 86140B Front and Rear Panels ...
Page 12: ...1 5 Getting Started Product Overview ...
Page 14: ...1 7 Getting Started Product Overview Figure 1 2 Display Annotations ...
Page 28: ...1 21 Getting Started The Softkey Panels ...
Page 33: ...1 26 Getting Started The Softkey Panels The Systems Menus continued ...
Page 41: ...1 34 Getting Started Product Options and Accessories ...
Page 79: ...2 38 Using the Instrument Analyzer Operating Modes ...
Page 80: ...3 Function Reference ...
Page 239: ...6 2 Maintenance Changing the Printer Paper Changing the Printer Paper ...