6-6
Maintenance
Printer Head Cleaning Procedure
C A U T I O N
Avoid dropping the coin or screwdriver, used to open the printer door, into the
printer assembly.
4
Lift up the paper latch as shown in the following diagram, and remove the
paper.
5
Unscrew the retaining screw that secures the sheet-metal cover that protects
the printer head from electrostatic discharge. Slide the sheet-metal cover
towards the retaining screw and then lift it straight up to remove.
6
Lift the printer head lever to the vertical position. Then, tilt the lever towards
the instrument’s rear panel to rotate the printer head up.