3-60
Function Reference
Perform Calibration
Perform Calibration
Performs a power or wavelength calibration after the setup has been specified
by the user.
Key Path
System > Calibration > Power Cal Setup > Perform Calibration
Remote
Commands
CALibration:WAVelength:INTernal
Pit Excursion (Marker Setup Panel)
Sets the pit excursion value for the marker search routines.
The pit excursion value is used to determine whether or not a local minimum
in the trace is to be considered a pit.
For marker search functions, a signal pit is defined as having a fall and then a
rise in the displayed response of at least the pit excursion value. Reducing the
pit excursion to values less than 3 dB may cause the marker-pitting functions
to identify noise spikes as pits. To reduce the noise floor variance to a value
less than the pit excursion rate, try changing the reference level at the top of
the graticule and increasing sensitivity.
Key Path
Markers > Marker Setup > Pit Excursion
Related Functions
Marker Search Threshold
Remote
Commands
CALCulate[1|2|3|4|5|6]:MARKer[1|2|3|4]:PEXCursion:PIT
Pit excursion criteria
Summary of Contents for 86140B Series
Page 2: ...Agilent 86140B Series Optical Spectrum Analyzer User s Guide ...
Page 11: ...1 4 Getting Started Product Overview Agilent 86140B Front and Rear Panels ...
Page 12: ...1 5 Getting Started Product Overview ...
Page 14: ...1 7 Getting Started Product Overview Figure 1 2 Display Annotations ...
Page 28: ...1 21 Getting Started The Softkey Panels ...
Page 33: ...1 26 Getting Started The Softkey Panels The Systems Menus continued ...
Page 41: ...1 34 Getting Started Product Options and Accessories ...
Page 79: ...2 38 Using the Instrument Analyzer Operating Modes ...
Page 80: ...3 Function Reference ...
Page 239: ...6 2 Maintenance Changing the Printer Paper Changing the Printer Paper ...