7-17
Specifications and Regulatory Information
Specifications
Table 7-14. Option 002 White Light Source
Agilent 86140B, 86141B, 86142B, 86146B only
Wavelength
a
a. Filtered below 850 nm.
900 nm to 1700 nm
Minimum Output Power Spectral Density
b
900 to 1600 nm
900 to 1600 nm (typical)
1600 to 1700 nm
b. With applied input fiber 9/125
µ
m.
–67 dBm/nm (0.2 nW/nm)
–64 dBm/nm (0.4 nW/nm)
–70 dBm/nm (0.1 nW/nm)
Minimum Output Power Spectral Density
c
d
50/125
µ
m fiber
62.5/125
µ
m fiber
c. Characteristic
d. Includes power in full numerical aperture of fiber.
–50 dBm/nm (10 nW/nm)
–46 dBm/nm (25 nW/nm)
Output Stability
±
0.02 dB over 10 minutes
Lamp Lifetime, Mean Time Between Failures
>5000 hours