7-7
Specifications and Regulatory Information
Specifications
Table 7-3. Amplitude Specifications
Amplitude
Agilent 8614xB
a
Sensitivity
b
600-750 nm
c
,
d
750-900 nm
900-1250 nm
1250-1610 nm
1610-1700 nm
e
–60 dBm
–75 dBm
–75 dBm
–90 dBm
–80 dBm
Maximum Measurement
Power
f
,
g
1525-1700 nm
600-1000 nm
1000-1525 nm
+15 dBm per channel,
+30 dBm total
+15 dBm per channel,
+30 dBm total
+12 dBm per channel,
+30 dBm total
Maximum Safe Power
Total safe power
Total power within any 10 nm
portion of the spectrum
+30 dBm
+23 dBm
Absolute Accuracy
h
at –20 dBm, 1310 nm/1550 nm
±
0.5 dB
Scale Fidelity
i
Autorange off
±
0.05 dB
Autorange on
±
0.07 dB
Display Scale
(log scale)
0.01-20 dB/DIV.
–120 to +90 dBm
Amplitude Stability
(1310 nm, 1550 nm)
1 minute
j
15 minutes
±
0.01 dB
±
0.02 dB
Summary of Contents for 86140B Series
Page 2: ...Agilent 86140B Series Optical Spectrum Analyzer User s Guide ...
Page 11: ...1 4 Getting Started Product Overview Agilent 86140B Front and Rear Panels ...
Page 12: ...1 5 Getting Started Product Overview ...
Page 14: ...1 7 Getting Started Product Overview Figure 1 2 Display Annotations ...
Page 28: ...1 21 Getting Started The Softkey Panels ...
Page 33: ...1 26 Getting Started The Softkey Panels The Systems Menus continued ...
Page 41: ...1 34 Getting Started Product Options and Accessories ...
Page 79: ...2 38 Using the Instrument Analyzer Operating Modes ...
Page 80: ...3 Function Reference ...
Page 239: ...6 2 Maintenance Changing the Printer Paper Changing the Printer Paper ...