OSA Errors
20005 desc
=
“Trans-Impedance calibration data is invalid”
help = “Factory calibration data for the trans-impedance amplifier is invalid.
Please record the hexadecimal number listed with the error and cycle power. If
the error persists, contact the nearest Agilent Technologies Instrument support
center for assistance. In the U.S., call (800) 403-0801. See the
Agilent 86140B
Series Users Guide
for a listing of the Agilent sales and service offices.”
20006 desc
=
“Monochromator flatness calibration data is invalid”
help = “Factory flatness data for the monochromator is invalid. Please record the
hexadecimal number listed with the error and cycle power. If the error persists,
contact the nearest Agilent Technologies Instrument support center for
assistance. In the U.S., call (800) 403-0801. See the
Agilent 86140B Series Users
Guide
for a listing of the Agilent sales and service offices.“
20007 desc
=
“Sweep Timeout”
help = “A sweep was started but did not finish in the expected amount of time.
The trace data acquired during this sweep may not be valid. Try taking another
sweep. If the error persists, contact the nearest Agilent Technologies Instrument
support center for assistance. In the U.S., call (800) 403-0801. See the
Agilent 86140B Series Users Guide
for a listing of the Agilent sales and service
offices.”
21999 desc
=
“The Error list has overflowed”
help = “The Error list has overflowed. The last entries received have been
deleted.”
Table 5-9. OSA Errors (2 of 2)
Error Number
Error Description [description/explanation/examples]
Summary of Contents for 86140B Series
Page 2: ...Agilent 86140B Series Optical Spectrum Analyzer User s Guide ...
Page 11: ...1 4 Getting Started Product Overview Agilent 86140B Front and Rear Panels ...
Page 12: ...1 5 Getting Started Product Overview ...
Page 14: ...1 7 Getting Started Product Overview Figure 1 2 Display Annotations ...
Page 28: ...1 21 Getting Started The Softkey Panels ...
Page 33: ...1 26 Getting Started The Softkey Panels The Systems Menus continued ...
Page 41: ...1 34 Getting Started Product Options and Accessories ...
Page 79: ...2 38 Using the Instrument Analyzer Operating Modes ...
Page 80: ...3 Function Reference ...
Page 239: ...6 2 Maintenance Changing the Printer Paper Changing the Printer Paper ...