3-104
Function Reference
Wavelength Units
Wavelength Units
Sets the display wavelength units to nm,
µ
m, or Ang.
Angtrom (Ang) is a unit of measurement of wavelength of light and other radi-
ation equal to one ten-thousandth of a micron or one hundred-millionth of a
centimeter.
Key Path
Wavelength > Wavelength Setup > Wavelength Units
Zero Now
Causes the auto zero function to run immediately. This allows you to manually
zero the instrument when the auto zero function is off.
In Auto Zero mode, the instrument compensates for temperature-related cur-
rent drift between each sweep. Although this “zeroing” increases amplitude
accuracy, it also increases the time between sweeps. If amplitude accuracy is
not critical to your measurement, turn off zeroing to improve throughput.
The instrument performs a more complete zeroing when the instrument is
first turned on. You can run this more complete routine anytime by pressing
the Zero Now softkey.
Key Path
System > More System Functions > Service Menu > Adv Service Functions
> Zero Now
Remote
Commands
CALibration:ZERO:AUTO ONCE
Summary of Contents for 86140B Series
Page 2: ...Agilent 86140B Series Optical Spectrum Analyzer User s Guide ...
Page 11: ...1 4 Getting Started Product Overview Agilent 86140B Front and Rear Panels ...
Page 12: ...1 5 Getting Started Product Overview ...
Page 14: ...1 7 Getting Started Product Overview Figure 1 2 Display Annotations ...
Page 28: ...1 21 Getting Started The Softkey Panels ...
Page 33: ...1 26 Getting Started The Softkey Panels The Systems Menus continued ...
Page 41: ...1 34 Getting Started Product Options and Accessories ...
Page 79: ...2 38 Using the Instrument Analyzer Operating Modes ...
Page 80: ...3 Function Reference ...
Page 239: ...6 2 Maintenance Changing the Printer Paper Changing the Printer Paper ...