Normal/Delta Marker Interpolation (Marker Setup
Panel)
Turns the normal/delta marker interpolation on or off. When on, the normal/
delta markers will be placed at the exact wavelength or frequency value
selected if it is within the trace range. The marker will linearly interpolate
between two trace data points. When off, the marker will be placed on the
nearest trace data point.
This setting enables interpolation for all four markers when in normal or delta
mode.
Key Path
Markers > Marker Setup > Normal/Delta Marker Interpolation
Related Functions
Active Marker, Delta Marker
Remote
Commands
CALCulate[1|2|3|4|5|6]:MARKer[1|2|3|4]:INTerpolation
Normal Marker Units (Marker Setup Panel)
Sets the marker X-axis readout for frequency or wavelength when the instru-
ment is in a non-zero span. The X-axis default units is nm (nanometers). This
setting controls only the normal marker X-axis readout and the delta refer-
ence readout. This setting controls all four normal markers. The bandwidth
and delta offset markers have their own setting.
Key Path
Markers > Marker Setup > Normal Marker Units
Related Functions
Active Marker
Remote
Commands
CALCulate[1|2|3|4|5|6]:MARKer[1|2|3|4]:X:READout
Summary of Contents for 86140B Series
Page 2: ...Agilent 86140B Series Optical Spectrum Analyzer User s Guide ...
Page 11: ...1 4 Getting Started Product Overview Agilent 86140B Front and Rear Panels ...
Page 12: ...1 5 Getting Started Product Overview ...
Page 14: ...1 7 Getting Started Product Overview Figure 1 2 Display Annotations ...
Page 28: ...1 21 Getting Started The Softkey Panels ...
Page 33: ...1 26 Getting Started The Softkey Panels The Systems Menus continued ...
Page 41: ...1 34 Getting Started Product Options and Accessories ...
Page 79: ...2 38 Using the Instrument Analyzer Operating Modes ...
Page 80: ...3 Function Reference ...
Page 239: ...6 2 Maintenance Changing the Printer Paper Changing the Printer Paper ...