Status Listings
OSA Warnings
5032–5043 desc
=
“I/O Error”
help “An unexpected error occurred during the I/O operation. Please try the
operation again. If the operation involves the floppy disk drive, try a different
floppy disk. If the error persists, please make a note of the error number and
contact the nearest Agilent Technologies Instrument support center for
assistance. In the U.S., call (800) 403-0801. See the
Agilent 86140B Series Users
Guide
for a listing of the Agilent sales and service offices.
5044 desc
=
“Please cycle power to synchronize system time”
help = “The system time clock has been set backwards. Due to internal system
requirements it is necessary to cycle power before continuing. Please turn
power off and back on again. This must be done to ensure proper system
operation.”
5045 desc
=
“Sweep limit markers too close together”
help = “The sweep limit function is on and the line markers are too close
together. The OSA cannot limit the sweep range to such a small value. The
current sweep may be using a wider range than requested. Move the line
markers farther apart or turn off sweep limit to resolve this problem.”
5046 desc
=
“AutoAlign cannot find input signal at marker”
help = “The auto-align procedure could not find a usable input signal. If there is
an active marker auto-align will attempt to align at the marker wavelength
without searching for a signal peak. This error can also occur if the input signal
amplitude is very low or if the OSA has received a large mechanical shock or if
the input signal is removed after the auto-align procedure has begun. Be sure a
signal is connected to the optical input. If there is an active marker make sure it
is placed on the input signal or turn off all markers.”
5047 desc
=
“Signal disappeared during AutoAlign”
help = “The auto-align procedure did not complete. The input signal disappeared
during the auto-align procedure. Make sure the input signal is connected to the
optical input and try again.”
5048 desc
=
“Wrong trace X axis units for peak to center”
help = “The peak-to-center function was requested. This function will only
execute if the active trace X axis has units of meters (um nm pm). Change the
active trace to one with an X axis in meters or re-measure the active trace in a
non-zero span.”
Table 5-6. OSA Warnings (4 of 12)
Error Number
Error Description [description/explanation/examples]
Summary of Contents for 86140B Series
Page 2: ...Agilent 86140B Series Optical Spectrum Analyzer User s Guide ...
Page 11: ...1 4 Getting Started Product Overview Agilent 86140B Front and Rear Panels ...
Page 12: ...1 5 Getting Started Product Overview ...
Page 14: ...1 7 Getting Started Product Overview Figure 1 2 Display Annotations ...
Page 28: ...1 21 Getting Started The Softkey Panels ...
Page 33: ...1 26 Getting Started The Softkey Panels The Systems Menus continued ...
Page 41: ...1 34 Getting Started Product Options and Accessories ...
Page 79: ...2 38 Using the Instrument Analyzer Operating Modes ...
Page 80: ...3 Function Reference ...
Page 239: ...6 2 Maintenance Changing the Printer Paper Changing the Printer Paper ...