3-32
Function Reference
Hold A...F None Min Max (trace)
Hold A...F None Min Max (trace)
Hold Max compares the current amplitude value of each point on the active
trace in the current sweep to the corresponding point detected during the
previous sweep, then displays the maximum value.
Hold Min compares the current amplitude value of each point on a trace in the
current sweep to the corresponding point detected during the previous
sweep, then displays the minimum value.
Hold None turns the Hold function off.
Key Path
Traces > Hold None Min Max
Remote
Commands
CALCulate[1|2|3|4|5|6]:MAXimum:STATe
CALCulate[1|2|3|4|5|6]:MINimum:STATe
Integrate Limit
Calculates the power between Wavelength Marker 1 and Wavelength Marker 2
when the trace integration function is on.
Once wavelength markers have been turned on, the total power integration
marker search, and wavelength sweep may be individually limited to the wave-
length marker range.
Key Path
Markers > More Marker Functions > Line Marker Menu > Advanced Line
Mkr Functions > Integrate Limit
Related Functions
Sweep Limit, Search Limit
Remote
Commands
CALCulate:TPOWer:IRANge:LOWer
CALCulate:TPOWer:IRANge:UPPer
Summary of Contents for 86140B Series
Page 2: ...Agilent 86140B Series Optical Spectrum Analyzer User s Guide ...
Page 11: ...1 4 Getting Started Product Overview Agilent 86140B Front and Rear Panels ...
Page 12: ...1 5 Getting Started Product Overview ...
Page 14: ...1 7 Getting Started Product Overview Figure 1 2 Display Annotations ...
Page 28: ...1 21 Getting Started The Softkey Panels ...
Page 33: ...1 26 Getting Started The Softkey Panels The Systems Menus continued ...
Page 41: ...1 34 Getting Started Product Options and Accessories ...
Page 79: ...2 38 Using the Instrument Analyzer Operating Modes ...
Page 80: ...3 Function Reference ...
Page 239: ...6 2 Maintenance Changing the Printer Paper Changing the Printer Paper ...