Troubleshooting Tasks
8
Troubleshooting
99
To Verify That the NPD Bead Is Ignited
To verify that the bead is ignited, look through the vent hole on
the detector lid to see if the bead is glowing orange.
The NPD output is selected by the operator as part of the adjust
offset process and generally is between 5.0 and 50.0 pA.
WA R N I N G
Hot exhaust! Detector exhaust is hot and can cause burns.