Troubleshooting Tasks
8
Troubleshooting
91
•
Flow-measuring insert (19301-60660)
•
Electronic flowmeter calibrated for the gas and flow rates
of concern
2
Set the bead voltage to 0.0 V.
3
Cool the NPD to 100 °C.
4
Remove the bead and store it carefully until re-installation.
5
Insert the NPD flowmeter adapter tool into the NPD
collector.
6
Attach the flow-measuring insert to the NPD flowmeter
adapter tool.
7
Place the flowmeter tubing over the flow-measuring insert to
begin measuring flows.