GC Not Ready Symptoms
4
Troubleshooting
61
A Gas Exceeds Pressure Setpoint or Flow
If a gas exceeds its pressure or flow setpoint, do the following:
If using a split/splitless inlet, PTV inlet, or volatiles interface:
•
Decrease the split ratio.
•
Replace the split vent filter.
•
Verify that the correct liner is selected (for split/splitless and
PTV inlets).
•
Check the gold seal for contamination (for split/splitless
inlet).
If using an FID or NPD:
•
Check for a plugged jet.
Valves:
•
Check for a misaligned rotor.