Chromatographic Symptoms
3
Troubleshooting
27
Baseline Rise During Oven Temperature Program
•
Inspect the column for bleed.
•
Check for leaks/oxygen in carrier gas supply.
•
Check gas supply oxygen trap indicator or date.
•
Make solvent blank runs to evaluate baseline without sample.
•
Make “no injection” blank runs (remove the syringe from the
injector and start a run) to evaluate baseline without solvent.
•
Check for contamination. (See Contamination or Carryover.)
•
Consider the effect of column film thickness on bleed.
•
Check for leaks at the column fittings. (See “Checking for
Leaks” .)
•
Prepare and use a column compensation profile.