Shutdown Symptoms
5
Troubleshooting
71
Hydrogen Shutdown
Hydrogen gas may be used as a carrier or as fuel for some
detectors. When mixed with air, hydrogen can form explosive
mixtures.
The GC monitors inlet and auxiliary gas streams. If a stream
shuts down because it is unable to reach its flow or pressure
setpoint and if that stream is configured to use hydrogen, the
GC assumes that a leak has occurred and shuts down
components to create a safe state. The GC:
•
Closes the carrier supply valve to the inlet and closes and
turns off both pressure and flow controls. When viewed,
these parameters will flash
Off
.
•
Opens the split vent valves in the split/splitless and PTV
inlets.
•
Turns off the oven heater and fan and opens the oven flaps.
•
Turns off all heaters (including any devices connected to the
auxiliary heater controls, such as valve box heaters and
transfer line heaters). When viewed, these parameters will
flash
Off
.
•
Sounds an alarm.
To recover from this state:
1
Fix the cause of the shutdown:
•
Replace the inlet septum.
•
Replace the inlet O-ring.
•
Check for broken column.
•
Check the supply pressure.
•
Check the system for leaks. See Checking for Leaks.
2
Press the key for the device that initiated the shutdown.
Scroll to the pneumatic parameter that is flashing
Off
, then
press [
On]
or [
Off]
. For example, if the front inlet ran out of
carrier gas, press [
Front Inlet
], scroll to the pressure or flow
parameter, then press [
On
].
WA R N I N G
The GC cannot detect leaks in the detector gas streams. For this
reason, it is vital that the column fittings of the FID, NPD, and any
other detectors that use hydrogen always be connected to a
column or have a cap or plug installed and that hydrogen streams
be configured so that the GC is aware of them.
Summary of Contents for 6890N
Page 1: ...Agilent Technologies Agilent 6890N Gas Chromatograph Troubleshooting ...
Page 20: ...20 Troubleshooting 2 Hardware Symptoms ...
Page 76: ...76 Troubleshooting 6 Nonfunctional GC Symptoms ...
Page 88: ...88 Troubleshooting 8 Troubleshooting Tasks ...
Page 102: ...102 Troubleshooting 8 Troubleshooting Tasks ...