Troubleshooting
8
Troubleshooting Tasks
To Measure NPD Leakage Current
1
Load the analytical method.
2
Set the
NPD Adjust Offset
to
Off
and the
Bead Voltage
to
0.00 V
.
•
Leave the NPD at operating temperature.
•
Leave flows on or off.
3
Press
[Front Detector]
or
[Back Detector]
, then scroll to
Output
.
4
Verify that the output (leakage current) is stable and
< 1.0 pA.
5
The output should slowly drop toward 0.0 pA, and should
stabilize in the
tenths
of a picoamp. Current > 2.0pA
indicates a problem.