Chromatographic Symptoms
3
Troubleshooting
29
Peak Tailing
The figure below shows an example of tailing peaks. When
troubleshooting tailing peaks, consider:
•
Which peaks are tailing?
•
Are the tailing peaks active compounds, all compounds, or
are there trends (such as early eluters or late eluters)?
•
Check the column for severe contamination.
•
Consider the column stationary phase (active column).
•
Verify that the column was cut and installed properly.
•
Consider the type of adapter, liner, and inlet seal being used.
One or all of these may be contaminated or active.
•
Check adapters (if installed) and liner for solid particles.
•
For capillary splitless injection, consider compatibility
between the solvent and column.
•
Verify that the injection technique is adequate.
•
Verify the inlet temperature.
•
Check for dead volume in the system. Check for correct
column installation at both ends.
•
Inspect any transfer lines for cold spots.
NPD
For NPD, do the following:
•
Verify that you are using the correct bead for the sample
being run. If you are analyzing phosphorus, install a black
bead. White beads can cause peak tailing when phosphorus
is being analyzed.
•
Verify that the correct jet is installed. Use an extended jet.
•
Replace the ceramic insulators.