Chromatographic Symptoms
Peak Fronting
The figure below shows examples of the three types of peaks:
symmetric, fronting, and overloaded.
If peak fronting or overloading occurs, try the following:
•
Verify that the injection volume is appropriate.
•
Ensure that the column is installed properly.
•
Verify that the appropriate injection technique is being used.
•
If using capillary splitless injection, consider the compound
solubility in the injection solvent.
•
Change the solvent.
•
Use a retention gap.
•
Check purity of sample solvent.
Symmetrical
Fronting
Overloaded