Troubleshooting Tasks
8
Troubleshooting
89
To Measure a Detector Flow
Measuring FID, TCD, uECD, and FPD flows
1
Gather the following:
•
Appropriate flowmeter adapter tube (can be found in the
GC ship kit)
•
Electronic flowmeter calibrated for the gas and flow rates
of concern
2
Set the oven temperature to ambient (35 °C).
3
Turn off the column flow and pressure.
4
Turn off (where applicable): the FID flame, FPD flame, and
TCD filament.
5
Cool the detector.
6
Connect the appropriate adapter to the detector exhaust.
A rubber adapter tube attaches directly to a uECD, or TCD
exhaust vent.
A separate adapter is supplied for the FID. Insert the adapter
into the detector exhaust vent as far as possible. You will feel
resistance as the adapter O-ring is forced into the detector
exhaust vent. Twist and push the adapter during insertion to
ensure a good seal.
For FPD, remove the vent tube assembly and use a 1/4-inch
tube adapter between the detector exhaust and the
flowmeter tubing.
WA R N I N G
Hydrogen (H
2
) is flammable and is an explosion hazard when
mixed with air in an enclosed space (for example, a flow meter).
Purge flowmeters with inert gas as needed. Always measure
gases individually. Always turn off detectors to prevent
flame/bead autoignition.
C A U T I O N
To avoid damaging the column, cool the oven before turning off the
column flow.
Summary of Contents for 6890N
Page 1: ...Agilent Technologies Agilent 6890N Gas Chromatograph Troubleshooting ...
Page 20: ...20 Troubleshooting 2 Hardware Symptoms ...
Page 76: ...76 Troubleshooting 6 Nonfunctional GC Symptoms ...
Page 88: ...88 Troubleshooting 8 Troubleshooting Tasks ...
Page 102: ...102 Troubleshooting 8 Troubleshooting Tasks ...