Hardware Symptoms
2
Troubleshooting
13
Vial Mishandled by ALS
Refer to the 7683B Automatic Liquid Sampler Installation,
Operation and Maintenance manual for additional information.
When you find a mishandled sample vial, do the following:
•
Check for folds or wrinkles in the crimp cap, especially near
the neck of the sample vial.
•
Use Agilent recommended sample vials.
•
Check sample labels (if applicable).
•
Check that they are the correct size.
•
Verify that the labels do not interfere with the gripper.
•
Check that the tray quadrants are clean and snapped into the
tray base.