Troubleshooting Tasks
8
Troubleshooting
95
To Verify FID Ignitor Function During Ignition Sequence
1
Remove the detector top cover.
2
Turn the FID flame
On.
3
Observe the ignitor plug though the FID chimney. The small
hole should glow during ignition sequence.
WA R N I N G
Keep body parts at a safe distance from the FID chimney while
performing this task. If using hydrogen, the FID flame will not be
visible.