Troubleshooting
2
Hardware Symptoms
Syringe Needle Bends During Injection into Inlet
Refer to the 7683B Automatic Liquid Sampler Installation,
Operation and Maintenance manual for additional information.
•
Check that the GC septum nut is not too tight.
•
Check that the syringe is installed correctly into the syringe
carriage.
•
Check that the needle support and guide are clean. Remove
any residue or septum deposits.
•
If using the cool on-column inlet, check that the correct
insert for the syringe is installed.
•
Check that you are using the proper syringe. The combined
length of the syringe barrel and needle should be
approximately 126.5 mm.
WA R N I N G
When troubleshooting the injector, keep your hands away from
the syringe needle. The needle is sharp and may contain
hazardous chemicals.