GC Not Ready Symptoms
4
Troubleshooting
67
NPD Adjust Offset Process Fails
•
Inspect the jet to see if it is clogged.
•
Measure the actual detector flows. (See To Measure a
Detector Flow.)
•
Check the condition of the bead. Replace if necessary.
•
Verify that the flow settings are correct.
•
Thoroughly leak check the whole system, especially the
detector column fitting. (See Checking for Leaks.)
•
Set the equilibration time to 0.0.