Chromatographic Symptoms
3
Troubleshooting
31
Sample Decomposition in Inlet/Missing Peaks
•
Lower the inlet temperature.
•
Check for air or water in the carrier gas; verify gas purity and
functionality of traps.
•
Verify that the liner is appropriate for the sample being run.
•
Perform complete inlet maintenance: Replace all consumable
parts and bake out the inlet.
•
Install a deactivated liner (SS, PP, and PTV inlets).
•
Check for leaks at the septum, liner, and column fittings. (See
“Checking for Leaks” .)
•
Install an Agilent Direct Connect liner.
•
Use a pulsed pressure method for quicker sample transfer to
column.
•
Bake out the inlet. See the following:
To Bakeout Contaminants from the Split/Splitless Inlet
To Bakeout Contaminants from the Purged Packed Inlet
To Bakeout Contaminants from the COC Inlet
To Bakeout Contaminants from the PTV Inlet
•