GC Not Ready Symptoms
Temperature Never Becomes Ready
To be considered ready, a temperature must be at setpoint ±1 °C
for 30 s. If a temperature never becomes ready, do the following:
•
Check for a missing insulation cup on an inlet or detector.
•
Check for a very large temperature difference between the
oven and inlet or detector.
•
Check for missing insulation around the inlet or detector.
•
If using a cool on-column with CryoBlast or a PTV inlet:
•
Check cryo coolant level.
•
Check if operating limits have been exceeded.