
Chromatographic Symptoms
3
Troubleshooting
39
FID Baseline Output Above 20 pA
•
Verify the purity of the carrier and detector gas supply.
•
Inspect the column for column bleed.
•
Check the gas supply trap indicators/dates and ensure that
the traps are not expended.
•
Verify that the detector was reassembled properly after
recent maintenance.
•
Inspect the detector for contamination.
•
Check that the FID leakage current is < 2.0 pA. (See “To
Measure FID Leakage Current” .)
Summary of Contents for 6890N
Page 1: ...Agilent Technologies Agilent 6890N Gas Chromatograph Troubleshooting ...
Page 20: ...20 Troubleshooting 2 Hardware Symptoms ...
Page 76: ...76 Troubleshooting 6 Nonfunctional GC Symptoms ...
Page 88: ...88 Troubleshooting 8 Troubleshooting Tasks ...
Page 102: ...102 Troubleshooting 8 Troubleshooting Tasks ...