Troubleshooting
3
Chromatographic Symptoms
FID Flame Goes Out During a Run and Attempts to Reignite
The following is an example chromatogram showing a flameout
from a large solvent peak.
After a flameout, the GC will try to ignite the flame three times.
If the FID flame goes out during a run, do the following:
•
See if an aromatic peak or water extinguished the flame.
•
Check for a plugged jet.
•
Verify that the gas flow settings are correct. Verify that
Lit
offset
is set appropriately.
If the FID flame attempts to reignite but is already lit, do the
following:
•
Verify that the FID
Lit offset
setting is appropriate for the run
(typically
<
2.0 pA).
•
Check to see if an aromatic peak or water extinguished the
flame.
•
Check for a partially plugged jet. Measure actual hydrogen,
air, and makeup flows at the detector. (See “To Measure a
Detector Flow” .)
•
Check for leaks at the detector column fitting. (See “Checking
for Leaks” .)
9
8
7
6
5
4
3
2
1
2
3
4
5 min
pA