70
Troubleshooting
5
Shutdown Symptoms
Column Shutdown
If a carrier gas (which can include an auxiliary flow module or
pneumatics control module) shuts down, the following happens:
•
The oven turns off to avoid column damage.
•
The oven flaps in the back of the oven open halfway.
•
The oven temperature setpoint flashes
Off
.
•
All flows for the column turn off. When viewed, their
parameters flash
Off
. For example, the septum purge and
column flows for a split/splitless inlet would turn off.
•
All other heaters turn off. When viewed, their temperature
parameters flash
Off
.
•
Attempts to turn on a shut-down zone fail with an error
message.
•
The instrument beeps.
To recover from this state.
1
Fix the cause of the shutdown.
•
Check for a broken column.
•
Replace the inlet septum.
•
Replace the inlet O-ring.
•
Check the supply pressure.
•
Check for leaks.
2
Press the key for the device that initiated the shutdown.
Scroll to the pneumatic parameter that is flashing
Off
, then
press [
On]
or [
Off]
.
For example, if the front inlet ran out of carrier gas, press
[
Front Inlet
], scroll to the pressure or flow parameter, then
press [
On
].