Troubleshooting
3
Chromatographic Symptoms
FPD Baseline Output High, > 20 pA
•
Check the supply gas purity.
•
Check all trap indicators and dates.
•
Check the detector for contamination.
•
Check for light leaks at the photomultiplier tube (PMT);
tighten the PMT if it is loose.
•
Perform complete inlet maintenance: Replace all consumable
parts and bake out the inlet.
•
Perform column maintenance: Bake out contaminants as
needed.