Troubleshooting Tasks
8
Troubleshooting
101
To Adjust the FPD Lit Offset
To adjust the FPD
Lit offset
:
1
Press
[Config]
.
2
Scroll to
Front detector
or
Back detector
(wherever the detector
is installed) and press
[Enter]
.
3
Scroll to
Lit offset
. With the
Lit offset
line highlighted, enter the
new parameter for the FPD (typical value is 2.0 pA), and
press
[Enter]
.