36
Troubleshooting
3
Chromatographic Symptoms
Check for the correct jet.
Low Peak Area or Height
•
If using an inlet in split mode, check the split ratio.
•
Check for leaks. (See “Checking for Leaks” .)
•
Check the inlet for contamination. (See “Contamination or
Carryover” .)
•
Check each column and verify that it was cut and installed
properly at each end.
•
Verify that the column type is correct.
•
Perform column maintenance: Bake out contaminants,
remove the contaminated length of column near the inlet,
and reverse and bake out the column as needed.
•
Verify that the liner type is appropriate for the sample.
•
Verify that the detector flow settings are correct.
•
Check the supply gas purity.
•
Check all trap indicators and dates.
•
Verify that the method parameters are correct.
•
Check sample stability.
•
Check configured syringe size. Some syringe sizes are
specified at half-capacity. If the maximum syringe volume is
marked at half-height on the barrel, not at the top of the
barrel, enter
twice
the labeled volume when configuring the
syringe size.
If using an FID:
•
Verify that the correct jet is installed.
•
Check for a dirty jet.
If using a uECD:
•
Replace the fused silica indented mixing liner.
•
Replace and reinstall column.
•
Clean the makeup gas adapter.
If using an NPD:
•
Check the detector for contamination.
•
Replace ceramic insulators.
•
Replace the bead.
If using an FPD:
•
Verify correct column installation.
Summary of Contents for 6890N
Page 1: ...Agilent Technologies Agilent 6890N Gas Chromatograph Troubleshooting ...
Page 20: ...20 Troubleshooting 2 Hardware Symptoms ...
Page 76: ...76 Troubleshooting 6 Nonfunctional GC Symptoms ...
Page 88: ...88 Troubleshooting 8 Troubleshooting Tasks ...
Page 102: ...102 Troubleshooting 8 Troubleshooting Tasks ...