Troubleshooting Tasks
8
Troubleshooting
97
To Measure FID Baseline Output
1
With the column installed, load your checkout method.
2
Set the oven temperature to 35 °C.
3
Press
[Front Detector]
or
[Back Detector]
, then scroll to
Output
.
4
When the flame is lit and the GC is ready, verify that the
output is stable and < 20 pA (this may take some time).
5
If the output is not stable or > 20 pA, the system or gas may
be contaminated. If this contamination is isolated to the
detector, bakeout the FID.